tive air humidity lower than 80 %.
7.1 AC voltage ranges
The input resistance is 2 MΩ in parallel 100 pF. The measurement reading is
obtained by rectification of average reading and displayed as actual value.
Measuring range Resolution
200 V
750 V
7.2 DC voltage ranges
The input resistance is 2 MΩ.
Measuring range Resolution
200 V
1000 V
7.3 AC current ranges
(Current-fork opening embraces AC wire).
Measuring range Resolution
200 A
Additional error in parallel current-conducting wire: < 0.08 A/ A.
7.4 Resistance measuring ranges
Overload protection for resistance measurements: 600 V
Measuring
range
200 Ω
2 kΩ
20 kΩ
200 kΩ
2 MΩ
20 MΩ
7.5 Diode and continuity test
The stated measuring accuracy applies to a range between 0.4 V and 0.8 V.
Overload protection: 600 V
The integrated buzzer sounds at a resistance R lower than 50 Ω.
Measuring
Resolution
range
1 mV
8.

Measuring with the BENNING CM 1-3

8.1 Preparing the measurement

Operate and store the BENNING CM 1-3 at the specified storage and operating
temperatures only! Do not permanently expose the device to sunlight.
-
Check stated nominal voltage and nominal current on the safety measuring
lines. Nominal voltage and current of the enclosed safety measuring lines
comply with the respective values of the BENNING CM 1-3.
-
Check insulation of the safety measuring lines. If the insulation is damaged,
the safety measuring lines must be replaced immediately.
-
Check the safety measuring lines for continuity. If the conductor in the
safety measuring line is interrupted, replace the safety measuring lines
immediately.
-
Before selecting another function by means of the rotary switch 2, discon-
05/ 2008
± (1.5 % of reading + 5 digit)
0.1 V
in frequency range 50 Hz - 500 Hz
± (1.5 % of reading + 5 digit)
1 V
in frequency range 50 Hz - 500 Hz
0.1 V
± (1.0 % of reading + 2 digit)
1 V
± (1.0 % of reading + 2 digit)
± (3.0 % of the measured value + 3 digits)
0.1 A
within the frequency range 50 Hz – 60 Hz
Resolution
0.1 Ω
± (1.0 % of the measured value+ 5 digits)
1 Ω
± (1.0 % of the measured value + 2 digits)
10 Ω
± (1.0 % of the measured value + 2 digits)
100 Ω
± (1.0 % of the measured value + 2 digits)
1 kΩ
± (1.0 % of the measured value + 2 digits)
10 kΩ
± (1.9 % of the measured value + 5 digits)
eff
Measuring accuracy
± (1.5 % of the measured value +
BENNING CM 1-3
Measurement accuracy
Measurement accuracy
Measurement accuracy
Measuring accuracy
0.05 V)
Overload protection
750 V
1000 DC voltage
750 V
1000 DC voltage
Overload protection
750 V
1000 DC voltage
750 V
1000 DC voltage
Overload protection
400 A
eff
Max. open-circuit
voltage
1.3 V
1.3 V
1.3 V
1.3 V
1.3 V
1.3 V
Max. measuring
open-circuit
current
1.5 mA
eff
eff
eff
eff
Max.
voltage
3.0 V
12