• Cleaning and disinfection should be performed using Clorox Healthcare Bleach Germicidal Wipes
at least once per week.
• The meter has a use life of 5 years and has been validated for 522 cleaning and 522 disinfection
cycles (the equivalent of 2 cycles per week for 5 years).
• The lancing device has a use life of 2 years and has been validated for 210 cleaning and
disinfection cycles (the equivalent of 2 cycles per week for 2 years).
• Cleaning and disinfecting the meter is important to prevent the spread of infectious diseases.
Cleaning and Disinfection
Cleaning is the physical removal of organic soil from the meter and lancing device surfaces.
Keeping the meter and lancing device clean helps ensure that they are working properly and that
no dirt gets in the devices. Cleaning allows for successful, subsequent disinfection.
Disinfection is a process that destroys pathogens, such as viruses and other microorganisms, on the
meter and lancing device surfaces. Disinfecting the meter and lancing device helps ensure that no
infection is passed on when you or others come in contact with the meter and lancing device. The
meter and lancing device should be cleaned with a separate bleach wipe prior to each disinfection.
Clean and disinfect your meter at least once a week. Clean and disinfect your lancing device when
visibly dirty (at least once a week).
Note: Squeeze bleach wipe to remove excess solution before wiping the meter or lancing device.
35
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22812-003_rev-A.indd 35-36
All manuals and user guides at all-guides.com
Cleaning
IMPORTANT: If you require assistance or if you notice any signs of deterioration on
the meter or lancing device (such as clouding or crazing on the display of the meter,
corroding or eroding of the plastic housing, or cracking of plastic housing or display) or
if the meter does not turn on or if the lancing device does not work after cleaning and
disinfection, discontinue use of the meter or lancing device and contact Customer Care
(1-888-522-5226).
Step 1
For cleaning, wipe all outside surfaces of the meter and lancing device thoroughly with
the bleach wipes until they are visibly clean. Clean the strip port, data port and battery
compartment with caution and do not force liquids into sensitive meter electronic
openings. Clean your meter at least once a week. Clean your lancing device when visibly
dirty (at least once a week).
Step 2
Allow the meter and lancing device to air dry completely before performing a blood
glucose test.
Step 3
When finished, thoroughly wash your hands with soap and water.