Freestyle Freedom Lite Manual Del Usuario página 22

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Disinfection
Step 1
For disinfection, wipe all outside surfaces of the meter and lancing device thoroughly
with the bleach wipe until they are wet in order to disinfect the meter and lancing
device. Clean the strip port, data port and battery compartment with caution and do not
force liquids into sensitive meter electronic openings.
Step 2
Allow the meter and lancing device surfaces to remain wet for 60 seconds (1 full minute).
After 1 full minute, immediately dry with drying towel to remove any residual moisture.
Disinfect your meter at least once a week. Disinfect your lancing device when visibly
dirty (at least once a week).
Step 3
Allow the meter and lancing device to air dry completely before performing a blood
glucose test.
Step 4
When finished, thoroughly wash your hands with soap and water.
37
Contact Customer Care Toll Free at 1-888-522-5226 www.MyFreeStyle.com
ART22812-003_rev-A.indd 37-38
All manuals and user guides at all-guides.com
Additional information about the risks for transmitting bloodborne pathogens to persons
undergoing fingerstick procedures for blood sampling can be found in the following references:
1. "FDA Public Health Notification: Use of Fingerstick Devices on More than One Person Poses Risk
2. "CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk
for Transmitting Bloodborne Pathogens: Initial Communication" (2010) http://www.fda.gov/
MedicalDevices/Safety/AlertsandNotices/ucm224025 .htm
for Transmitting Bloodborne Pathogens" (2010) http://www .cdc .gov/injectionsafety/Fingerstick-
DevicesBGM.html
38
11/10/16 11:02 AM

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido