FOR USE WITH
GAME READY
GRPRO
2.1 SYSTEM
®
MED4 ELITE™ SYSTEM
SIZING SPECIFICATIONS
• Size: One Size
ASSEMBLY
The Heat Exchanger
(PN 520502-03)
comes as part of a pre-assembled Wrap
Exchanger is removed to launder the Sleeve or for any other reason,
please use the following instructions to re-insert the Heat Exchanger
into the Sleeve.
Lay the Sleeve with the zipper side
1
down on a flat surface and remove
the chest straps (C), then turn the Sleeve
over so the zipper is facing up.
Lay the Heat Exchanger out flat,
2
separate from the outer sleeve.
Insert the Heat Exchanger into the Sleeve
by feeding the connector through the
opening (F) on the left shoulder. The blue
side of the Heat Exchanger should
touch the blue side of the Sleeve.
Be sure the Heat Exchanger is flat
3
inside the sleeve on the back and
shoulder areas. In order for the collar
to lay correctly, rotate the shoulder
wings back (A), then smooth the heat
exchanger inside the collar. Zip the
sleeve closed, and flip the Wrap over
with the shoulder wings facing forward. The
Wrap should resemble a vest with a high
collar. Attach the chest straps (C) to complete
assembly, orienting the logo on the chest
straps as shown.
REMOVAL OF HEAT EXCHANGER
1. Disconnect the Connector Hose from the Wrap
2. Unzip the zipper
3. Gently pull out the Heat Exchanger
fits inside the Sleeve
(PN 510502)
. If the Heat
(PN 590502-03)
WARNING
It is mandatory to fully read and understand your
System's User Manual before using the device.
and
Failure to follow operating instructions could result in
serious injury.
IMPORTANT
Read complete indications, contraindications,
cautions, and warning before using this product.
Keep this document for future reference.
BEFORE INITIAL USE
Prime the Wrap using the following steps:
• With the GRPro 2.1 Control
Unit off, attach the Connector
Hose to the Control Unit and
the Wrap
• Lay the Wrap open and flat next
to the Control Unit (not on the
body)
• Turn the System on and run for
2 minutes with No Pressure
STORAGE OF YOUR WRAP
Hang your Wrap on a wide hanger or lay flat. Do not fold or stack it, as
this could kink the fluid chamber and the Wrap will not work properly.
C-T SPINE
USE GUIDE
ENGLISH
• With the Med4 Elite Control
Unit off, attach the Connector
Hose to Patient 1 on Control
Unit and the Wrap
• Lay the Wrap open and flat next
to the Control Unit (not on the
body)
• Press the On/Off button above
the touchscreen
• Choose Patient 1 and run Cold
Therapy with No Pressure for 2