4-3 Indicator Reference Mark (Zero) Adjustment

4-4 Modify Alarm Set-Point

8. Tighten set screw.
9. Install reed switch holder onto isolation well.
10.Tighten screw on reed switch holder.
11.Connect wiring per appropriate wiring diagram, Figure 1-7, 1-8 or 1-9.
12.To adjust alarm set point refer to Section 4-4.
13.Replace cover.
This adjustment is performed at a no flow condition with the float at rest.
1. Completely stop fluid flow through the meter.
2. Loosen the screw on the scale.
3. Align the "Ref" (zero) mark on the scale with the line on the indicator.
4. Tighten the scale screw.
1. Unscrew and remove alarm housing cover.
2. Loosen screw in reed switch holder.
3. Required alarm setting can be achieved by sliding reed switch holder
up or down on isolation well until desired alarm position is reached.
4. Tighten screw in reed switch holder securely.
5. Replace cover.
Section 4 Maintenance
MT36xx Series
4-3