4. OPERATING INSTRUCTIONS

4.1.

START-UP

Once the above points have been controlled, start up the MSC by pressing the On/Off button
(Annex A3, pos. 1). After entering the password correctly (by default the password is
0000) or immediately if this function has been disabled, the fans will automatically connect.
Ensure that the aspiration area and air exhaust outlet are free (Annex A2, pos. 8).
The electronic system of regulation by microprocessor via a thermoanemometer probe,
laminar down flow and exhaust airflow, needs a few seconds to stabilise during connection in
order to begin to work normally.
The cabinet can only be used by qualified staff who meets the correct work techniques in the
laboratory. If these techniques are not fulfilled, faulty working of the equipment may
happen.
WARNING
The GS certification does not allow the deactivation of the password enquiry.
4.2.

FRONT PANEL POSITIONS

By means of the handles "A", the front glass can be moved from working position to closed
position or vice versa. To move the glass just slide down or up the glass.
FROM WORKING POSITION
In case it is needed to clean the glass by the inner part (Annex A5.2) or if it is necessary to
place in the working area any big product or device, the glass-frontal group may be
separated from the cabinet, providing full access to both the inner glass area and the
working area. All needed precautions should be taken as the user protection decreases
during that operation.
UM-BIOIIADVANCE-EN-1302
All manuals and user guides at all-guides.com
FROM CLOSED POSITION
13