METRIA M51 Benchtop Conductivity/TDS Meter
1.4
Place the electrode into the conductivity standard solution, the meter automatically shows current calibration standard (e.g.,
1413µS/cm).
1.5
Press the
key, the default calibration value begins flashing.
Enter
Press the ◄ or ► key to set the calibration value, press the
1.6
completed, ensure that displayed value matches your calibration standard.
1.7
Press the
key, the meter begins the calibration, the Calibration icon will continuously flashing. Wait for the reading to stabilize,
Enter
the display
will show END. Single point calibration is completed.
Multi-point calibration
2.1
Ensure that you have selected 2 or 3 points calibration in the setup menu.
2.2
Repeat steps 1.2 to 1.7 above, when the first calibration point is completed, the display will show "CAL2". The meter prompts you to
continue with second point calibration.
key to confirm and move to next digit. When the setting is
Enter