6.2 Operation

After the power cable and light guide are properly connected, turn on the illuminator by pressing the
lamp power switch located on the front panel. The power indicator light within the switch should
illuminate, the fan will start, and the touchscreen will display a boot screen for three seconds. After
boot, the main operating screen will display.
Boot Screen

6.2.1 Fault Condition Notices

6.2.1.1
Engage Lamp
6.2.1.2
Close Door
LIT-218 Sunoptic Surgical
Rev. F
Refer to Section 8.1 – Lamp Replacement.
Disconnect power. Open the lamp door. The lamp module
positioning lever should be pressed fully down (horizontal) to
properly engage the lamp and fully raised (vertical) to
release it. Raise the lever and firmly push the lamp along the
black module shroud to ensure the lamp module is fully
seated on the power connectors. Lower the lamp
engagement lever, from vertical to horizontal to re-engage
the lamp module, ensuring that the lamp module rises
slightly.
ACTION: Close the lamp door completely.
NOTE: Fan will run with lamp door open.
®
Page 10 of 146