Auto power off: The meter will automatically power off to save power if there is no operation
for 15 min. You can wake it up by pressing any button or restart it after turning the switch to
OFF.
Press and hold the SELECT button in off state and then turn on the meter again to disable
the auto power off function. Restart the meter after shut it down to resume this function.
Buzzer: When any button is pressed or the function switch is turned, if it is valid, the buzzer
will make one beep (about 0.25s). The buzzer will also beep intermittently to indicate the over
range during the voltage or current measurement.
Low battery detection: The battery voltage will be automatically detected as long as the
meter is on. If it is lower than 2.6V, the LCD will display the "
Low battery shutdown function: When the battery voltage is lower than 2.5V, the LCD
displays the "
" symbol, the "Lo.bt" interface appears and lasts for about 10s, the buzzer
makes consecutive beeps 3 times, and then the meter automatically shuts down (no interface is
displayed).
TESTING IN CAT III MEASUREMENT LOCATIONS
Ensure the test lead shield pressed firmly in place. Failure to use the CAT III shield
increases arc-flash risk.
TESTING IN CAT II MEASUREMENT LOCATIONS
CAT III shields may be removed for CAT II locations. This will allow testing on recessed
conductors such as standard wall outlets.
PROBE USAGE
" symbol.
21