GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
GUÍA Y DECLARACIÓN DE ORTOTECSA – INMUNIDAD ELECTROMAGNÉTICA
The infrared lamp is intended for use in an electromagnetic environment specified below. The customer or user of the infrared lamp
should assure that it is used in such an environment
Immunity
IEC 60601 test
Test
Level
Ensayo de
Nivel de ensayo de
inmunidad
la norma
Conducted RF
3 v
rms
IEC 61000-4-6
150 kHz to 80 MHz
RF conducida
3 v/m
Radiated RF
80 MHz to 2,5 GHz
IEC 61000-4-3
RF radiada
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To access the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the infrared lamp is used exceeds the applicable RF compliance level above, the XXXX
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocation the infrared lamp.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Ortotecsa,s.l. C/Peña de Francia,8
Tlf.:+34918719132
USER AND MAINTENANCE MANUAL
Compliance
Level
Nivel de conformidad
3 V
rms
3 V/m 80 MHz to 1
GHz (due to the
testing laboratory
limitations)
(debido a las
limitaciones de
laboratorio)
28500 Arganda del Rey (Madrid), España.
Fax: +34918719323
email:
Rev. 4. Date: Ag 28, 2020
Electromagnetic Environment Guidance
Entorno electromagnético, guía
Portable and mobile RF communications equipment should be
used no closer to any part of the infrared lamp, including
cables, tan the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Los equipos móviles y comunicaciones de RF no se deben
usar más cerca de cualquier parte de la lámpara, incluyendo
los cables, que la distancia de separación recomendada a la
frecuencia del transmisor.
Recommended separation distance:
Distancia de separación recomendada:
d = 1,17 √P 150 kHz to 80 MHz
d =1,17√P 80 MHz to 800 MHz
d=2,33√P 800 MHz to 2,5 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in
meters (m).
Donde p es la maxima potencia de salida asignada de
transmisor en vatios (w) conforme al fabricante del transmisor
y d es la distancia de separación recomendada en metros.
Field strengths from fixed RF transmitters, as determined by
an electromagnetic site survey a, should be less than the
compliance level in each frequency range.
Las intensidades de campo desde el transmisor fijo de RF,
según se determina por un estudio electromagnético de lugar,
debe ser menor que el nivel de conformidad en cada rango de
frecuencia.
Interference may occur in the vicinity of the equipment
marked with the following symbol:
La interferencia puede ocurrir en la vecindad del equipo
marcado con el siguiente símbolo:
info@ortotecsa.com
www.ortotecsa.com
b
pag 14 of 15