Timer
G.18900-A and G.18900-A/1
Formula to calculate the dosing techniques applicable energy on the body
Dose ( J/cm2) x Surface(cm2)
Treatment time(s) = ---------------------------------------
W (power average)
WARNING!
Sometimes it iwill be applied anti-inflammatories, analgesics, etc. To take advance on the hyperemia and
dilation of the pores produced by the heat thus favoring absorption. In any case will be done before, but
after the application of IR, to prevent possible overheating of the area. It will be removed all metal items
such as jewelry, etc., Because the rapid heating of the metal can cause burns on contact area.
Install the equipment in order to be easy to disconnect in case of failure.
Dont manipulate or open the lamp: it is not allowed except by the personnel of the manufacturer.
BULB REPLACEMENT
In case you need to replace the bulb for having faulty or damaged, it should be performed with a lamp
230V/115V (SEE LABEL LAMP) and the thread diameter and bulb that corresponds to the defective.
Recommended Replacement bulbs are:
BULB OSTRAM THERATTHERM 230 V.
-
BULB PHILIPHS BR125 IR 250W.
-
BULB PHILIPHS PAR38 IR 150W E27
-
If the bulb broke unplug the lamp and let it cool.
Beware of glass rests that could remain in the body of the lamp and carefully using pliers (or similar tool)
remove the threaded portion of the broken bulb before replacing it with the new bulb.
The breakage and/or replacement of the bulb is not part of the warranty for this product.
Dont try to change the bulb without previously disconnect the lamp from the plug.