2. "CH1" will appear on the display. Press the "CHANNEL" button until the display
again shows the message corresponding to the desired channel (order of
messages displayed:
"CH1" -> "CH2" -> "CH3" -> "IN" -> "CH1").
3. After selecting the desired channel, hold down the "ALERT" button and after a
moment press it again.
4. The following parameters will be displayed sequentially:
- upper limit of temperature value (orange display backlight),
- Lower limit of temperature value (red highlighted),
- upper limit of the humidity value (orange highlight of the display),
- lower limit of humidity value (red backlight),
- output.
5. Use the "UP/MAX/MIN" and "DOWN/C/F" buttons to adjust the limits.
A single press will change -/+ 0.1
in the value, it is recommended to hold down the desired button (0.8
change per second).
6. When the measurement exceeds the set limit, the unit will illuminate with a
given color (orange or red), give a warning signal for 5 seconds and show
warning icons on the display.
The illumination in the warning color and the warning signal will appear every
minute until the measurement is no longer over the limit.
To manually silence the warning signal, press any button.
7. The default set limits are:
- upper for indoor temperature:
or 1 %. In order to make significant changes
o
or 8 %
o