Hold Measurement - Keysight 34450A Guia Del Usuario

Multímetro de 5 1/2 dígitos
Ocultar thumbs Ver también para 34450A:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Hold measurement

The hold feature allows you to capture and hold a stable reading on the front
panel display.
When a stable reading is detected, the multimeter emits a beep (if the buzzer is
enabled in the utility menu) and holds the reading on the primary display.
Primary display
(stable hold
reading)
Figure 2-2
When enabled, the hold operation turns on the hold annunciator and begins
evaluating readings using the rules described below:
Primary Display = Reading
The decision to update a new reading value in the primary display is based upon
the box-car moving statistics of the present reading and the three previous
readings:
Max (Reading
Min (Reading
Keysight 34450A User's Guide
>
Press to enable hold measurement
Typical hold function display
>
Press to disable hold measurement
Accessing hold measurement
IF Max()
N
Reading
Reading
N
N
1
Reading
Reading
N
N
1
Functions and Features
Secondary display
(present measurement
reading)
Min()
0.1% × Reading
Reading
)
N
2
N
3
Reading
)
N
2
N
3
2
1
2
3
N
53

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido