Changing the Battery:
This device is supplied with one lithium cell (CR2032 x 1).
1. Open the battery cover: Insert a pointed object into the battery cover pick hole. At the same time, use
thumb to remove battery cover. (See Figure 1)
2. Flip the battery out with a small screw driver (See Figure 2)
3. Insert the new battery under the metal hook on the left side and press the right side of the battery
down until you hear a "click". (See Figure 3)
4. Replace the battery cover.
The positive (+) side Up and the negative (-) side pointed Down.
Figure 1
Specifications:
- Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
- Storage temperature Range: It should be stored at room temperature between -20~+50°C, RH ≤85%
Transportation temperature shall be less than 70°C, RH ≤95%
- Atmospheric pressure: 800~1013 hPa
- Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC), IEC/EN60601-1(Safe-
ty) standards, ISO10993, RoHS.
- Accuracy: ±0.2°C (0.4°F) within 35~42°C (95~107.6°F), ±0.3°C (0.5°F) for other range.
- This thermometer is an adjusted mode thermometer that converts the ear temperature to display its
"oral equivalent" (according to the result of the clinical evaluation to get the offset value)
- Enclosure Rating: IP22
- Dimensions: 149.1 x 42.5 x 54.9 mm
- Weight: 79.7 grams including battery
- Battery life: around 3,000 continuous readings.
- Product life: 4 years
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using infrared thermometer.
This is not an AP or APG product.
NOTE: The thermometer is calibrated at the time of manufacture. If at any time you question the accu-
racy of temperature measurements, please contact the dealers or nearest service address.
ENGLISH
Figure 2
8
Figure 3