3.5.1 Emergency stop Eagle 620
3.5.2 Emergency stop Eagle 625
In emergencies, and during assembly and disassembly, the
emergency stop is activated by pressing the red button on the
control box.
Push on the button and turn in the direction of the arrows. If the
button is fully pressed to the back turn back in the opposite
direction of the arrows.
The emergency stop is deactivated by turning the red button in
the direction of the arrows.
3.5.3 Emergency lowering Eagle 620
3.5.4 Emergency lowering Eagle 625
WARNING: Risk of injury – Adjust the manual emergency
L
lowering for the weight of the patient.
Emergency lowering if the power fails or the batteries are empty
can be done using the red pull button  at the lower end of the
motor. This emergency lowering is factory-set for a patient weight
of 75 kg.
Standard calibration: 3000N, 16 mm/s to lower the patiënt.
Please note that the manual emergency lowering is only possible
when the patient is sitting in the patient hoist.
In emergencies, and during assembly and disassembly, the
emergency stop is activated by pressing the red button on the
right of the control box.
Push on the button and turn in the direction of the arrows. If the
button is fully pressed to the back turn back in the opposite
direction of the arrows.
The emergency stop is deactivated by turning the red button in
the direction of the arrows.
If the manual operation mechanism should be damaged, it is
possible to lower the cantilever arm from the control box.
The arm moves down for as long as the bottom button on the
control box panel is pressed.
12
Eagle 620, Eagle 625
2022-04
Using the patient hoist