Service and Repair
Send the MedStream Control Unit, universal battery charger, and USB cable
for service or repair to:
Codman Neuro Sciences Sàrl
Rue Girardet 29
CH 2400 Le Locle
Switzerland
Always include a repair purchase order and a written description of the
problem.
End of Useful Life
Dispose of the equipment in accordance with local ordinances.
Control Unit Technical Specifications
Internal Power Supply
(non-removable)
Charger max. voltage
Charger max. current
Approximate stand-alone operating
time (includes 48 radio-frequency
communications with the pump
within this period of time)
Maximum battery recharge time
(The Control Unit is operational with
depleted battery while connected to
the charger.)
Weight
Dimensions
206102-001-D.indb 7
206102-001-D.indb 7
Lithium-ion rechargeable battery
Nominal voltage: 11.1 V
Minimum capacity: 1850 mAh
Battery configuration: 3S1P
Active current limitation: 4.65 A ±
1.05 A
12.6 V DC
1.2 A DC
8 h
5 h
750 g
130 mm W x 60 mm H x 110 mm D
Control Unit Environmental Conditions
Operating Conditions
Temperature Range:
Humidity Range:
Pressure Range:
Transport & Storage Conditions
Temperature Range:
Humidity Range:
Pressure Range:
Warranty
Codman & Shurtleff, Inc., warrants that this medical device is free from
defects in both materials and workmanship for one (1) year from the date of
purchase. Any other express or implied warranties, including warranties
of merchantability or fitness, are hereby disclaimed. Suitability for use
of this medical device for any particular surgical procedure should
be determined by the user in conformance with the manufacturer's
instructions for use. There are no warranties that extend beyond the
description on the face hereof.
™ MedStream is a trademark of Codman & Shurtleff, Inc.
10°C to 40°C
Relative Humidity 30% to 75%
non-condensing
Atmospheric Pressure
700 hPa – 1060 hPa
Transport: –20°C to +50°C
Storage: +5°C to +30°C
Relative Humidity 10% to 85%
non-condensing
Atmospheric Pressure
500 hPa – 1060 hPa
7
9/15/2014 6:26:37 PM
9/15/2014 6:26:37 PM