Guidance and manufacturer's declaration – electromagnetic immunity
Triangle is intended for use in the electromagnetic environment specified below. The customer
or the user of the device should assure that they are used in such an environment.
Immunity Test
Electrostatic
± 8 kV contact
discharge (ESD)
± 2 kV, ± 4 kV,
IEC 61000-4-2
± 8 kV, ± 15 kV air
± 2 kV for power
Electrical fast
supply lines
transient / burst
± 1 kV for input /
IEC 61000-4-4
output lines
± 1 kV differential
mode
Surge
IEC 61000-4-5
±2 kV common
mode
< 5% U
(> 95% dip in U
for 0,5 cycle.
Voltage dips, short
40% U
interruptions and
(60% dip in U
voltage variations
5 cycles.
on power supply
70% U
input lines
(30% dip in U
IEC 61000-4-11
25 cycles.
<5% U
(> 95% dip in U
for 5 s.
Power frequency
(50/60 Hz)
30 A/m
magnetic field
IEC 61000-4-8
Note:
(1)
A lock or reboot of the device with no permanent damage is acceptable
(2)
U
is the a.c. main voltage prior to application of the test level.
T
IEC 60601
Compliance Level
test level
± 8 kV contact
± 2 kV, ± 4 kV,
± 8 kV, ± 15 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV differential
mode
±2 kV common mode
(2)
< 5% U
T
)
(> 95% dip in U
T
0,5 cycle.
40% U
T
) for
(60% dip in U
T
cycles.
70% U
T
) for
(30% dip in U
T
25 cycles.
<5% U
T
)
(> 95% dip in U
T
5 s.
30 A/m
Electromagnetic
Environment
Floors should be wood,
concrete or ceramic tile.
(1)
If floors are covered
with synthetic material,
(1)
the relative humidity
should be at least 30%
Mains power quality
should be that of a
professional healthcare
facility environment.
Mains power quality
should be that of a
professional healthcare
facility environment.
(2)
Mains power quality
T
should be that of a
) for
T
professional healthcare
facility environment. If
the user of Triangle
T
) for 5
requires continued
T
operation during a
power mains
T
interruption, it is
) for
T
recommended that
Triangle be powered
with an uninterruptible
T
) for
power supply or a
T
battery.
Power frequency
magnetic field should be
at levels characteristic
of a professional
healthcare facility
environment.
Guidance
37
EN