Injector manifolds
with screw thread

Installation

Unscrew the upper part  from the holder.
Fit a silicone adapter (ADS)  in its place which has the same dia-
meter as the transmission instrument being reprocessed.
 Damage to filter plates
Fitting a transmission instrument into a silicone adapter that is too
big can cause damage to the filter plate. Damage to filter plates
may affect their ability to effectively filter out dirt particles from the
wash water, leading to blockages in the channels of the transmis-
sion instruments.
Only use silicone adapters (ADS) with the same diameter as the
transmission instrument being reprocessed.
An A 801 adapter is required.
Remove an injector nozzle from the injector manifold and replace it
with the A 801  adapter.
Screw the lower part of the  holder into the adapter.
Insert a filter plate  into the lower part.
 Do not insert a filter plate into the holder if the injector manifold
has been fitted with an A 800 tubular filter.
Screw the upper part into the lower part.