Reprocessing transmission instruments

Before loading check that the silicone adapters are present and
correctly fitted.
Wipe away any deposits from the area around the silicone adapters
that may be present depending on local water conditions using a
damp cloth.
Check that internal sections of transmission instruments are free of
blockages.
Fit the instruments into their holders using the respective silicone
adapters.
After reprocessing, the insides of the instruments must be dried
with sterile compressed air before being cleaned and sterilised as
appropriate in accordance with the manufacturer's instructions.
Please observe national regulations.
Before using transmission instruments again following repro-
cessing, a function check must be carried out, e.g. by spraying into
a basin, to ensure they are clear.