Disturbance emission tests
Description of the tests
Radiated RF emission test
Conducted disturbance voltage on AC
Flicker
Immunity tests
Description of the tests
Immunity against radiated
RF disturbances
Immunity against conducted
RF disturbances on AC power
and patient cables
Immunity against EFT signals
Immunity against surge voltages
Immunity against voltage dips
Immunity against short voltage
interruptions
Immunity against electrostatic
discharges (ESD)
Flicker measurement was omitted because no fl icker was observed.
User manual * DMR-01/A05/V08 * 2022-08-07 * Nosiboo Pro Nasal Aspirator * Page: 15/52
Limits and test levels of
the related Standard
FCC Part 15, 2007 30-3000 MHz class B
FCC Part 15, 2007 0.15-30 MHz class B
EN 60601-1-2:2014
EN 61000-3-3:2009
Limits and test levels of
the related Standard
EN 60601-1-2:2014
3V/m 0.08-2.5GHz
Modulation: 1 kHz, 80 % AM
EN 60601-1-2:2014
3 Veff 0.15 – 80 MHz
Modulation: 1 kHz, 80 % AM
EN 60601-1-2:2014
±2 kV between L, N and ref. ground
EN 60601-1-2:2014
±1 kV between L and N
EN 60601-1-2:2014
100 %,60 %,30 %
0.5, 5, 25 periods
EN 60601-1-2:2014
100 % for 5 periods
EN 60601-1-2:2014
±8 kV air, ±6 kV contact
Result
Passed
Passed
NA
Result
Passed
Evaluation: operation in compliance
with the specifi cation
Passed
Evaluation: operation in compliance
with the specifi cation
Passed
Evaluation: operation in compliance
with the specifi cation
Passed
Evaluation: operation in compliance
with the specifi cation
Passed
Evaluation: operation in compliance
with the specifi cation
Passed
Evaluation: operation in compliance
with the specifi cation
Passed
Evaluation: operation in compliance
with the specifi cation
ENG