Blood pressure classification in adults
Systolic
(mmHg)
Severe hypertension
Moderate hypertension
Mild hypertension
Normal-high
Normal
6.10. Description of technical alarms
The monitor will display "Hi" or "Lo" as a technical alarm on the screen without delay if the
determined blood pressure (systolic or diastolic) is outside the nominal range specified in the
SPECIFICATIONS section. In this case, you should consult a physician or check if the operation
complies with the instructions.
The technical alarm condition (out of nominal range) is set at the factory and cannot be adjusted
or disabled. This alarm condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm does not latch and no reset is required. The signal displayed on the LCD
will disappear automatically after approximately 8 seconds.
DE
PT NL
SWE
Diastolic
(mmHg)
WHO definitions and classification of blood
pressure levels
PL
DK
Classification
PSS (mmHg)
Pressure
Optimal
<120
Normal
120-129
Normal-high
130-139
Hypertension
140-159
grade 1
Hypertension
160-179
grade 2
Hypertension
≥180
grade 3
Blood Pressure Monitor
PSD
Colour
(mmHg)
indicator
<80
green
80-84
green
85-89
green
90-99
yellow
100-109
orange
≥110
red