6.11 Troubleshooting
Problem
LCD display
shows an abnor-
mal result
LCD shows
battery symbol
LCD shows "Er 0"
LCD shows"Er 1"
LCD shows "Er 2"
LCD shows "Er 3"
LCD shows "Er 4"
LCD shows "Er 5"
LCD shows "Er 6"
LCD shows "Er 7"
LCD shows "Er 8"
LCD shows "Er A"
DE
PT NL
SWE
Possible cause
The position of the cuff is not
correct or has not been properly
adjusted.
Body posture during the test
was not correct.
Talking,
arm
movements,
during the test
Irregular heartbeat (arrhythmias)
Low battery
Pressure system is unstable
before measurement
Does
not
detect
pressure
Does
not
detect
pressure
Pneumatic system blocked or
cuff too tight during inflation
Leaks in the pneumatic system
or cuff is too tight during inflation
cuff pressure above 300 mmHg
More than 160 seconds with cuff
pressure above 15 mmHg
Internal memory error
Device parameter check error
Pressure sensor parameter error
PL
DK
Blood Pressure Monitor
Solution
Place the cuff correctly and try
again.
Review the "BODY POSTURE
DURING MEASUREMENT" section
in the instructions and try again.
or
body
Try again when you are calm and
nervousness
do not talk or move during the
test.
This electronic blood pressure
monitor is not suitable for use by
persons with severe arrhythmias.
Replace the batteries
systolic
Do not move and try again
diastolic
Apply the cuff correctly and try
again.
Measure again after five minutes.
If the monitor is still not abnormal,
please
distributor or the factory.
contact
your
local
25