153
ELECTROMAGNETIC COMPATIBILITY (EMC)
• This device should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this
device should be observed to verify normal operation in the configuration in which it will be used.
• Use of accessories other than those specified or provided by the manufacturer of this device could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no
closer than 30 cm (12 inches) to any part of the device, including cables specified by the manufacturer. Otherwise, degradation
of the performance of this equipment could result.
Declaration - electromagnetic emission
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of device should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Table 1
Compliance
Electromagnetic environment - guidance
The device uses RF energy only for its
internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause
any interference in nearby electronic
equipment.
The device is suitable for domestic
establishment and in establishment directly
Class B
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Not applicable
Not applicable
Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below.
The customer or the user of device should assure that it is used in such an environment.
Immunity test
IEC 60601test level
Electrostatic
±8 kV contact
discharge (ESD)
±2 kV, ±4 kV, ±8 kV,
IEC 61000-4-2
±15 kV air
± 2 kV for
Electrical fast
power supply lines
transient/burst
± 1 kV for input/output
IEC 61000-4-4
lines
± 0.5kV,
Surge
± 1 kV line(s) to lines
IEC 61000-4-5
± 0.5kV, ± 1 kV, ± 2 kV
line(s) to earth
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
Voltage dips, short interruptions
and voltage variations on
0 % UT; 1 cycle
power supply input lines
and70 % UT;
IEC 61000-4-11
25/30 cycles
Single phase: at 0°
0 % UT;
250/300 cycles
Power frequency
(50/60 Hz) magnetic field IEC
30 A/m
61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Table 2
Electromagnetic environment -
Compliance level
guidance
Floors should be wood, concrete
±8 kV contact
or ceramic tie. If floors are covered
±2 kV, ±4 kV, ±8 kV,
with synthetic material, the relative
±15 kV air
humidity should be at least 30 %.
Mains power quality should be that
Not applicable
of a typical commercial or hospital
environment.
Mains power quality should be that
Not applicable
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the device
requires continued operation during
Not applicable
power mains interruptions, it is
recommended that the device be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
30 A/m
of a typical location in a typical
commercial or hospital environment.