Guidance and manufacturer's declaration –electromagnetic immunity
®
The PRONEB
ULTRA II is intended for use in the electromagnetic environment specified below. The customer or the user of
®
the PRONEB
ULTRA II should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic discharge
±6 kV contact
(ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV for input/output lines
IEC 61000-4-4
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV common mode
Voltage dips, short
<5 % U
T
interruptions and
(>95 % dip in U
voltage variations on
for 0,5 cycle
power supply input
lines
40 % U
T
(60 % dip in U
T
IEC 61000-4-11
for 5 cycles
70 % U
T
(30 % dip in U
T
for 25 cycles
<5 % U
T
(>95 % dip in U
for 5 sec
Power frequency
(50/60 Hz) magnetic
field
3 A / m
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
English
Compliance level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV differential mode
±2 kV common mode
<5 % U
T
)
(>95 % dip in U
)
T
T
for 0,5 cycle
40 % U
T
)
(60 % dip in U
)
T
for 5 cycles
70 % U
T
)
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
)
(>95 % dip in U
)
T
T
for 5 sec
3 A / m
Electromagnetic environment
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
®
of the PRONEB
ULTRA II
requires continued operation
during power mains
interruptions, it is recommended
®
that the PRONEB
ULTRA II be
powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.