3. Install the cover on the battery pack again:
a. The cover fits into the battery pack only one way. The opening next to the USB mark must
align with the USB connector in the battery pack.
b. Tighten the cover until there is no gap between the bottom O-ring and the battery pack
housing.
4. If the battery pack was removed from the pressure housing, use the banana plug sockets on the
bottom plate of the battery pack to verify that the battery voltage is approximately 19.5 volts.
5. Install the battery pack in the pressure housing if it was removed.
a. Make sure there is no water on any surface.
b. The O-rings must be pristine. Apply a small quantity of Parker Super O-Lube on any new O-
rings, then carefully install the new ones.
6. Carefully put the end flange in to the pressure housing and tighten into position. Use a wrench if
necessary.
Section 4 General information
This product can expose the user to chemicals with silica, crystalline (airborne particles of respirable size), which
is known to the State of California to cause cancer and birth defects or other reproductive harm. For more
information, go to www.P65Warnings.ca.gov.