Technical Information
The controller can answer with the following response types:
Table 12
Response Type
Logic
Numeric
Alphanumeric
ACK
NACK
Unknown Window
Data Type Error
Out of Range
Win Disabled
170
Length
Value
1 byte
-
6 bytes
-
Up to 14 bytes
-
1 byte
(0x6)
1 byte
(0x15)
1 byte
(0x32)
1 byte
(0x33)
1 byte
(0x34)
1 byte
(0x35)
Agilent TwisTorr Medium-TMP Rack Controller User Manual
Description
After a read instruction of a logic
window.
After a read instruction of a numeric
window.
After a read instruction of an
alphanumeric window.
The command execution has been
successfully completed.
The command execution has been
failed.
The specified window in the command
is not a valid window.
The data type specified in the
command (Logic, Numeric or
Alphanumeric) is not accorded with
the specified window.
The value expressed during a write
command is out of range value of the
specified window.
The specified window is Read Only or
temporarily disabled (for example) you
can't write the Soft Start when the
Pump is running).