• Wear appropriate personal protective equipment during sterilization and lubrication process. This should include eye
protection and gloves.
• Do not use ultrasonic bath or cold sterile/disinfectant chemicals to clean handpiece.
• Do not autoclave handpieces together with other instruments.
• Do not use dry heat sterilization.
• Keep handpiece in sterilization pouch until ready to use.
• Do not cool handpiece manually as damage may occur.
• Do not autoclave at temperatures that exceed 275º F (135º C).
1. Remove bur, then wipe external surface with dry paper towel to remove any excess lubricant and debris. Place paper towel in
bio-hazard bin when finished.
2. Place handpiece in sterilization pouch and seal. Bag handpieces individually; no more than 1 per pouch.
3. Sterilize in a gravity type autoclave for a minimum of 15 minutes or in a pre-vacuum autoclave for a minimum of 4 minutes at
270° F (132° C). Let autoclave cycle completely. Do not remove before handpiece is completely cool and pouch is dry.