Cleaning and Lubrication with EVERclean Handpiece Maintenance Machine
Handpiece maintenance
Note: Cleaning and lubrication is required before each sterilization cycle.
1. Remove bur.
2. Detach handpiece from delivery system.
3. Attach handpiece to appropriate connector in EVERclean Maintenance Machine (5028208U0).
4. Run maintenance cycle. See EVERclean instructions for further information.
5. Remove handpiece from EVERclean Maintenance Machine after cycle completes.
6. Wipe excess cleaner/lubricant from handpiece with dry, lint-free cloth.
Chuck maintenance
Note: It is recommended that the push-button chuck be cleaned every 3 - 4 maintenance cycles.
1. Remove bur.
2. Place chuck opening against tapered end of chuck cleaning nozzle.
3. Press start. Keep handpiece in place until beep sounds indicating that cleaning cycle is completed. See EVERclean instructions
for additional information.
8 EN