SOLAR-02
26
5.5.2
USE OF THE UNIT
1. Switch on the unit and press the SET key. The unit shows the screen beside
in which sensor 1 is selected
2. Use the arrow keys ▲and ▼ to select the desired type of sensor considering
the unit programming (see § 5.2)
3. Confirm selection with the SET key to go back to the measuring mode
4. Connect the unit SOLAR-02 to the master instrument by means of the USB
cable. The symbol "USB" is shown on the display while the unit is detected
by the master instrument
5. By activating a testing/recording on the master instrument, the unit
SOLAR-02 waits for the synchronization, showing the message "HoLd" on
the display and the indication of the waiting time necessary before being
able to activate the testing procedures (see example in the screen beside)
6. Once the waiting phase (HOLD) has ended, the message "Recording..." will
be displayed. It is now possible to disconnect the USB cable from the unit
SOLAR-02 and take the unit near the PV modules. Possible hyphens "- - -"
indicate that no irradiation and temperature probes are connected to the
unit SOLAR-02 yet
7. Place the Irradiance probe (Irr-P or Irr-S2 – see Table 1) parallel to PV
modules surface and connect the output connector to SOLAR-02 PYRA/
CELL input
8. Using a tape fix the temperature probe (Tmp – see Table 1) on the backside
of the module and connect the output connector to SOLAR-02 TEMP input
9. Upon reaching an irradiation value higher that the threshold set by the
MASTER instrument, the unit SOLAR-02 shows the message "READY" on
the display
10.
Disconnect the irradiation and temperature probes, connect the unit
to the master instrument again by means of the USB cable
11. Upon detecting the control to stop testing/recording from the master
instrument, the unit stops working, shows on the display the message
"SEnd" and sends the data to the master instrument to determine the final
result of the testing
5.6
USING THE SOLAR-02 WITH MASTER INSTRUMENT TYPE "M-RF"
For the definitions of the instrument types, see Table 1. For the use, functions
and activation of a testing/recording on the master instrument, please see
the user manual of the master instrument itself.
5.6.1
PRELIMINARY CHECK
For the "M-RF" master instruments the enable of internal RF module IS
NECESSARY. Check that the parameter which enable/disable the RF module
should be corrected set at the first usage of the unit in the herewith way:
By press and hold the FUNC/▲ key while switching on the meter by
key. The message "rF
" is displayed for a few seconds. If the message "rF
ON
displayed switch off the unit and repeat the operation. Afterwards the initial
screen is displayed by the meter. The setting of "RF" parameter is maintained
also during the replace of batteries
Rel. 4.00 - 22/11/22