Table 4-1
Default factory settings and available setting options for each feature (continued)
Menu
Feature
BAUD
DATA BIT
PARITY
4
ECHO
PRINT
REVISION
S/N
V-ALERT
5
M-INITIAL
SMOOTH
DEFAULT
6
BATTERY
DC FILTER
Keysight U1253B User's and Service Guide
Default factory
Available setting options
setting
9600
Baud rate for remote communication with a PC (remote control). Available
options: 2400, 4800, 9600, and 19200.
8
Data bit length for remote communication with a PC. Available options: 8 bits or
7 bits (stop bit is always 1 bit).
NONE
Parity bit for remote communication with a PC. Available options: NONE, ODD,
or EVEN.
OFF
Return of characters to PC in remote communication. Available options: ON or
OFF.
OFF
Prints measured data to a PC in remote communication. Available options: ON
or OFF.
NN.NN
Revision number. Editing is disabled.
NNNNNNNN
The last 8 digits of the serial number will be indicated. Editing is disabled.
OFF
Audible alert tone for voltage measurement.
– To enable this function, select an overvoltage value within the range of
1 V to 1010 V.
– To disable this function, set all digits to zero ("OFF" will be indicated).
FACTORY
Initial measurement functions. Available options: FACTORY or USER.
NORMAL
Refresh rate for primary display readings. Available options: FAST, NORMAL, or
SLOW.
NO
Select YES, then press
to its default factory settings.
7.2 V
Battery type used for the multimeter. Available options: 7.2 V or 8.4 V.
OFF
Filter for DC voltage or DC current measurement. Available options: OFF or ON.
Changing the Default Settings
for longer than 1 second to reset the multimeter
4