On the rare occasion of a fault causing the cuff to
remain fully inflated during measurement, open the cuff
immediately. Prolonged high pressure (cuff pressure
300mmHg or constant pressure 15mmHg for more than
3 minutes) applied to the arm may lead to an ecchymosis.
Always be sure that operation of the device does not
result in prolonged impairment of patient
blood circulation.
The device cannot be used with HF surgical equipment
at the same time.
CAUTION:
This SPHYGMOMANOMETER was clinically investigated
according to the requirements of ISO 81060-2:2013.
Besides providing inaccurate readings, the effects of
this device on the fetus are unknown. Too frequent and
consecutive measurements could cause disturbances in
blood circulation and injuries.
This unit is not suitable for continuous monitoring during
medical emergencies or operations. Such use will cause
the patient's arm and fingers to become anaesthetic,
swollen and even purple due to a lack of blood.
When not in use, store the device with the adapter in a dry
room and protect it against extreme moisture, heat, lint,
dust and direct sunlight. Never place any heavy objects on
the storage case.
6
This device may be used only for the purpose described in
this booklet. The manufacturer cannot be held liable for
damage caused by incorrect application.
This device contains sensitive components and must be
treated with caution. Observe the storage and operating
conditions described in this booklet.
The user should not have a skin temperature that exceeds
42.8°C (109°F). The temperature of the room in which this
device is used should not exceed 40°C (104°F).
This device is not AP/APG equipment and not suitable
for use in the presence of a flammable anesthetic mixture
with air or with oxygen or nitrous oxide.
WARNING:
Do not service or perform maintenance on
this ME equipment while in use.
The patient is an intended operator.
The patient can measure, transmit data and recharge
device under normal circumstances and maintain the
device and its accessories according to the user's manual.
To avoid measurement errors, please avoid conditions
with strong electromagnetic field radiated interference
signals or electrical fast transient/burst signals.
The blood pressure monitor, its adapter, and the cuff are
suitable for use within the patient environment. If you are
allergic to polyester, nylon or plastic, please don't use this
device. During use, the patient will be in contact with the cuff.