Descargar Imprimir esta página

SHIMADZU CLAM 2040 CL Manual De Instrucciones página 274

Publicidad

11 Solución de problemas
Código
de error
0x070291
0x070292
0x070293
0x0702A2
0x0702A3
0x0702A4
0x0702B2
0x0702B3
0x0702B4
0x0702B6
0x0702B7
0x0702B8
0x0702C2
0x0702C3
0x0702C4
0x0702FE
0x0702FF
232
CLAM™-2040 CL
Mensaje de error
An abnormality has occurred
in the sample probe vertical
driving motor.
An abnormality has occurred
in the sample probe rotating
motor.
An abnormality has occurred
in the sample syringe pump
motor.
Sample probe protocol error
(Sample dispense)
Sample probe protocol error
(Sample dispense)
Sample probe protocol error
(Sample dispense)
Sample probe protocol error
(Detergent suction)
Sample probe protocol error
(Detergent suction)
Sample probe protocol error
(Detergent suction)
Protocol command for
dispensing sample with
diluent cannot finish (time
over).
Protocol command for
dispensing sample with
diluent cannot start
(abnormal parameter).
Protocol command for
dispensing sample with
diluent cannot start
(exclusive control).
Sample probe protocol error
(Washing)
Sample probe protocol error
(Washing)
Sample probe protocol error
(Washing)
Abnormal data ROM
A system error occurred.
Estado de la
unidad
Acción de usuario necesaria
después del
error
STOP
Póngase en contacto con el representante de
STOP
Shimadzu.
STOP
STOP
STOP
STOP
STOP
STOP
STOP
Confirme que la combinación del software del
STOP
ordenador de la consola y la versión de
firmware sea la adecuada.
STOP
STOP
STOP
STOP
STOP
STOP
Póngase en contacto con el representante de
Shimadzu.
STOP

Publicidad