1. Normal working conditions
Environmental temperature: 15°C–30°C
Relative humidity: ≤70%
Power supply: 5V
Input power: 10W
2. Transportation and storage conditions
Temperature: -20°C–55°C
Relative humidity: ≤80%
3. Parameters
3.1 Basic parameters
Sample size: 1
Dimensions (mm) (length × width × height): 101 × 91 × 65
Weight (g): 210g
3.2 Performance parameters
Sample module operating temperature range: 30°C–75°C
Maximum heating rate of sample module: ≥ 10°C/min
Temperature control accuracy of sample module: ±0.5°C
Sample module temperature uniformity: ±1.5°C (after 5 mins)