4.2 Indicator light status:
Power indicator light flashing, temperature preparation
stage (measurement cannot be started)
When the power indicator light stops flashing and turns
blue, the preparation is complete and the test can be started.
The power indicator light will flash blue during the test.
When the positive indicator light is on, the sample is
determined to be positive.
When the negative indicator light is on, the sample is
determined to be negative, or the sample concentration
value is too low to be detected.
When the negative and positive indicator lights are both
on, this indicates invalid test result. It may also indicate
insufficient sample volume, suppressed reaction, sampling
error or sample contamination during the test process.
When the power indicator flashes red and the beeper
sounds simultaneously, this indicates that the device is in
error state.
- 12 -