Technical Specifications
The 4UHV Controller is for internal use only.
Table 1
Component
Input Voltage
Input Frequency
Input Power
Output HV Voltage
Output HV Short circuit current
Output HV Power max
Operating temperature
Storage temperature
Fuse
Voltage measurement
Current measurement
Weight
Compliance with:
Installation category
Max altitude
Pollution degree
Internal use only
Agilent 4UHV Ion Pump Controller User Manual
Technical Specifications
Description
100 - 240 Vac (+/-10 %)
50/60 Hz
600 VA
+/- 7000 Vdc +/- 5 %
40 mA (for 80 W channels) 100 mA
(for 200 W channels)
200 W or 80 W each channel
0 °C to 40 °C
-20 °C to +70 °C
5x20 T6.3A 250 V
Resolution 100 V
Measurement range: 10 nA to 100 mA,
logarithmic scale
Resolution : 2 digit of the mantissa
(maximum 10 nA)
Accuracy: +/- 6 %
6.5 kg
EN 61010: 2010 AMD 1 (2017): Safety
requirements for electrical equipment for
measurement control and laboratory use.
EN 61326-1 2013: Electrical equipment for
measurement control and laboratory use EMC
requirements
II
2000 m
2