Technical Information
Spare Pump
How to select the Ion pump
1
Press MODE for at least 3 seconds to enter the programming menu.
2
Press the key " – " to select HVx, where x corresponds to the HV channel.
3
Press the key ENTER.
4
Press the key " – " to select the pump.
5
Press the key ENTER.
6
Press the key ESC to exit.
7
Press the key MODE for 3 seconds to exit the programming menu, otherwise
after 5 seconds of inactivity, the unit exits form programming menu
automatically.
How to work with non Agilent pump
Select the pump SPARE, in pump selection menu (see above).
1
Press MODE for at least 3 seconds to enter the programming menu.
2
Press the key " – " to select HVx, where x corresponds to the HV channel.
3
Press the key ENTER.
4
Press the key " – " to select the pump SPARE.
5
Pressure values are required, at 100 nA, 10 µA, 10 mA and 400 mA.
6
Press the key MODE for 3 seconds to exit the programming menu, otherwise
after 5 seconds of inactivity, the unit exits form programming menu
automatically.
7
Make sure that the interlock is connected; refer to to the warning at the
"Fixing the 4UHV Ion Pump Controller" paragraph.