Technical Information
The controller can reply with the following types of response:
Table 10
Type
Logic
Numeric
Alphanumeric
ACK
NACK
Unknown Window
Data Type Error
Out of Range
Win Disabled
198
Length max
Value
-
6 bytes
-
48 bytes
-
1 byte
(0x6)
1 byte
(0x15)
1 byte
(0x32)
1 byte
(0x33)
1 byte
(0x35)
Description
After a read instruction of a logic window.
After a read instruction of a numeric window.
After a read instruction of a alphanumeric window.
Execution of the command has been successful.
Execution of the command has failed.
The window specified in the command is not a valid
window.
The data type specified in the command (Logic,
Numeric or Alphanumeric) is not in agreement with
the Window specified.
The value expressed during a write command is not
within the range value for the specified window.
The window specified is Read Only or is temporarily
disabled.
Agilent 4UHV Ion Pump Controller User Manual