Technical Information
Configuration Input Structure
Total buffer length: 40 bytes (9 bytes + 8 floating)
Table 23
Offset
Unit/Res.
(byte)
0
1
-
2
.
3
-
4
1 bit = 100V
5
1 bit = 100V
6
1 bit = 100V
7
1 bit = 100V
8
1 bit = 100V
Agilent 4UHV Ion Pump Controller User Manual
Size
Description
(byte)
B
→ Communication stetus between CPU and internal profibus card
7
B
→ Communication status (0= ok, 1= not good)
6
B
→ 1=remote, 2=local, 3=Profibus
1,0
1
B
→ HV 1 On/Off
7
B
→ HV 2 On/Off
6
B
→ HV 3 On/Off
5
B
→ HV 4 On/Off
4
B
→ HV1 protect mode On/Off
3
B
→ HV2 protect mode On/Off
2
B
→ HV3 protect mode On/Off
1
B
→ HV4 protect mode On/Off
0
1
B
→ HV 1 mode (0= Fixed, 1= Step)
7
B
→ HV 2 mode (0= Fixed, 1= Step)
6
B
→ HV 3 mode (0= Fixed, 1= Step)
5
B
→ HV 4 mode (0= Fixed, 1= Step)
4
B
→ Set Point A ( 0= not active, 1 = active)
3
B
→ Set Point B ( 0= not active, 1 = active)
2
B
→ Set Point C ( 0= not active, 1 = active)
1
B
→ Set Point D ( 0= not active, 1 = active)
0
1
B
→ HV 1 cable interlock
7
B
→ HV 2 cable interlock
6
B
→ HV 3 cable interlock
5
B
→ HV 4 cable interlock
4
B
→ HV 1 Remote on/off
3
B
→ HV 1 Remote on/off
2
B
→ HV 1 Remote on/off
1
B
→ HV 1 Remote on/off
0
1
Controller status ( 0= no error, 6 = error)
1
Channel 1, High voltage measured
1
Channel 2, High voltage measured
1
Channel 3, High voltage measured
1
Channel 4, High voltage measured
225