Technical Information
How to set the Analog signal
1
2
3
4
5
6
Connection of Input Remote
In order to activate the inputs remote (CH on/off, Ch protect, CH Step)
connect the pins as following.
Figure 57 Input
Agilent 4UHV Ion Pump Controller User Manual
Press MODE for at least 3 seconds to enter the programming menu.
Press ENTER to enter the GENERAL configuration menu.
Use the arrow to scroll through the menu.
Press ENTER to make changes.
Use the key " + " or " – " to change the value; use the arrow to move the square
through the line; it's possible to set the analog signal either proportional to
current or to pressure.
Press the key MODE for 3 seconds to exit the programming menu, otherwise after
5 seconds of inactivity, the unit exits form programming menu automatically.
+24V
4,7 KΩ
A
Controller
Input pin
Digital GND
Customer
device's
193