This manual is intended to explain the installation and functioning of the device.
Warnings: This symbol shall be used in the manual to indicate situations of neglect which
could cause injuries or material considerable damage.
Cautions: This symbol shall be used in the manual to indicate information about
important cautions to prevent undesired consequences.
Read this manual carefully before using the device.
Should you have any doubt regarding the contents of the manual herein, or the operation and
safety of the device, contact your local distributor or the nearest ECLERIS office.
2.1 WARNINGS
It is important to read this manual carefully before using the Microstar Microscope.
ECLERIS will not be held responsible for damage that can occur to the device due to not
following the instructions or not observing the following criteria:
To avoid risk of electric shock, this device must only be connected to an earth-protected
power supply.
Do not install the equipment next to heat sources, such as heaters.
The Microscope must be used in medical facilities with a proper electrical installation,
including a grounding in compliance with the regulations.
The microscope should be connected to the mains through the supplied cord or a
replacement with the same certifications.
The mains should be grounded with no adapter needed.
The sectioning of the electrical network is carried out by disconnecting the power cable, or
by removing the connector from the electrical outlet.
The device is connected to a ground protection through a ground conductor in the power
supply cable. This ground connection is essential for the safe use of the equipment
The device is provided with a connectable power cable that assures proper electrical
isolation. If it needs to be replaced, it should have the same features than the original.
Only connect the device to visualization devices that meet the IEC 60601-1 standard.
Device not appropriate to be used near or around flammable anesthetic gas mixture with air,
oxygen or nitrous oxide.
Do not look directly into the light output; the optical head has a very bright light output.
To move the microscope (during installation and intended use) verify light is off, in order to
avoid bright light on the face.
Do not insert objects into the holes. Do not place any container with liquids on top of the
device.
In case of interference with other electro-medical or electric equipment, it is recommended
increasing the distance between the device and the other equipment to minimize these
effects.
INMANUUML019 USER MANUAL - M. USO OM-200 V.2023.01 ENG-ESP US
USER MANUAL / MANUAL DE USUARIO – MICROSTAR
2. SAFETY INFORMATION
Page 5 of 64