ON CLEANING
Recommended Cleaning Chemicals
•
Isopropanol 70 %
•
Ethanol 70 %
•
Cidex
®
OPA
•
0.9 % NaCl solution
How to Use Cleaner
•
Prior to cleaning, turn off the detector and remove the power cable.
•
Soak a soft cloth in a recommended cleaner, then lightly rub the screen with no more than 1 N of force.
•
The cleaner could cause serious damage if it leaks inside the detector while cleaning.
•
Do not use benzene, thinner, acids or alkaline cleaners or other such solvents.
•
Cleaning guidelines for detector must only be carried out by medical professionals (doctors or nurses) and
must not be handled by patients.
GENERAL DESCRIPTION
Overview
This model is an x-ray imaging device, a system that can acquire and process X-ray images as digital images. It
utilizes amorphous silicon and a high-performance scintillator to ensure sharp high-definition image quality
with the resolution of 3.6 lp/mm and the pixel pitches of 140 um. This device is a flat panel based X-ray image
acquisition device. This device must be used in conjunction with an operating PC and an X-ray Generator. This
device can be used for digitizing and transferring X-ray images for radiological diagnosis. The data transmission
between the detector and PC can be enabled with a wired (cable).
Product Components
•
Detector:
•
System Controller:
- AC Power Cord for the System Controller
•
Cable
- Main Cable: Detector and System Controller link cable (Supply DC power, Ethernet data, control signals of
X-ray Generator)
- Trigger Cable: X-ray Generator to System Controller, transmit control signal between detector and X-ray
Generator.
- LAN Cable: System Controller to PC, exchanges Ethernet data between PC and detector.
•
CD:
•
Owner's Manual (Detector, System Controller), Inspection Report