To reflect the newly modified Bad Pixel map in the future calibration, you can upload it to the detector.
•
Press the [Bad Pixel Map Upload] button and select some of the generated map files ([BPM.raw], [Avgdark.raw], [Offset.raw], [Gain.raw]). Then, the file will be uploaded.
•
When the file is successfully uploaded, you can see it on the Log.
NOTE
•
If the [Bad Pixel Map] is uploaded, the existing map in Factory Calibration folder will be deleted. To keep the map, you need to first back it up to another path.
•
To select a file to be uploaded for [Bad Pixel Map] upload, all four [Bad Pixel Map] files must be in the path. ([BPM.raw], [Avgdark.raw], [Offset.raw], [Gain.raw])