No.
LED Indicator
2
Connecting the System Controller and the Detector A. This connector supply power (24 V
3
Connecting the System Controller and the Detector B. This connector supply power (24 V
System Controller supports 2 Detector connections. Usage is, one is for Bucky stand, the other is for table (bed). Generally, X-ray room of hospital installs 2 Detectors, Bucky stand and
table type, it's far more convenient and efficient working environment. These 2 Detectors are not operated simultaneously, System Controller selects the operating Detector by AWS
command.
4
Connects AC power cord.
5
Ethernet port to transmit image/command between the Detector and PC.
6
This is to synchronize the Detector and X-ray Generator.
7
System Controller power fuses are 4 A, 250 V to Type T fuse.
Power rating:
Status
2.1 A) to the Detector, transmits X-ray synchronization signals and Ethernet image data.
2.1 A) to the Detector, transmits X-ray synchronization signals and Ethernet image data.