Transport
If the measurement indicates that the measured air is
contaminated with oil (applies only to high-pressure systems):
8. Unscrew the sintered filter
of the pressure regulator.
9. Insert and tighten a new
sintered filter.
10. Purge the device with
clean air for at least 6
minutes.
11. In case of severe
contamination, remove
the measuring device.
Clean the individual
components (excluding
the metering valves) with soapy water.
12. Dry with oil-free and dry air.
13. Purge the metering valve
with oil-free and dry air
only in the direction of the
arrow.
14. Mount the measuring
device.
15. If necessary, clean the snapper container and tube holder
with clean water (only the bottom, outlet side) and dry.
40
7
Transport
Transport only in the original packaging
8
Storage
Store the product in its original packaging. Store all parts in the
supplied case without deforming in a dry, cool and dust-free
place. Protect from exposure to direct sunlight and heat.
Observe ISO 2230. Store tubes securely, inaccessible to
unauthorized persons.
9
Disposal
Dispose of tubes according to the instructions for use.
Dispose of the Impactor with normal household waste.
Dispose of the tips inside the tube opener according to the
applicable waste disposal regulations.