en Quick Setup Guide
InPro X1 HLS pH Sensor
InPro X1 HLS
pH Sensor
Quick Setup Guide
1
2
3
4
5
© 04 / 2023 METTLER TOLEDO
Printed in Switzerland
0389147_C_QSG_InPro_X1_18lang_100x297mm_2023_04.indd 5
0389147_C_QSG_InPro_X1_18lang_100x297mm_2023_04.indd 5
Intended Use
This sensor is a food-safe, Clean-in-Place (CIP) resistant
single-rod measuring cell for in-line measurement of pH
values in food and beverage applications. It meets the re-
quirements of 3A, EHEDG, EC1935 / 2004 and ATEX guide-
line 2014 / 34 / EU. Declaration of conformity: www.mt.com/
InProX1HLS.
Read the instruction manual before commissioning.
Only operate the electrode with original parts from METTLER
TOLEDO.
Only trained personnel must operate and maintain the sensor.
Safety and Cleaning Instructions
When handling or calibrating the sensor, personal protec-
tive equipment such as protective spectacles and gloves
should be worn.
To ensure hygienic operation, the sensor can be exposed to
CIP procedures. The maximum tested conditions are: 3 %
NaOH, 90 °C, 30 min; H
30 min; H
O, 90 °C, 30 min.
2
Information on the Sensor
InPro X1
ABC-Dxxx-Eyyy
X ... Y °C (Z °C)
pH X ... Y
Further information: Order number, identification and num-
ber of certificates, code for FM testing and serial number on
the sensor body.
Instructions
For more information on installation, commissioning, cal-
ibration, cleaning, storage, disposal, ordering and confor-
mity, see: www.mt.com/InProX1HLS.
• Download the 3A and EHEDG certificates;
• Download the EC1935 / 2004 declaration
of conformity for food contact materials;
• Download the instruction manual.
Environmental Protection
Waste electrical products should not be disposed of with
household waste. Please recycle where facilities exist. Check
with your local authority or retailer for recycling advice.
O, 90 °C, 30 min; 3 % HNO
2
Type designation
A
= Certifications
B
= Reference system
C
= Tip design
D
= Redox yes / no
xxx = Shaft materials
E
= Cable connection
yyy = Rod length in mm
Operating temperature range (for cleaning)
pH measurement range
en
5
, 45 °C,
3
InPro X1
30 389 147 C