Technical data
Mode of operation
Measuring range > 100 to
< 600 mbar
Input
Measured variable input
Measuring range for
Overload limit
gauge pressure
0 ... 100 mbar g
> -400 / < 400 mbar g
0 ... 160 mbar g
> -400 / < 400 mbar g
0 ... 250 mbar g
> -800 / < 1 000 mbar g
0 ... 400 mbar g
> -800 / < 1 000 mbar g
0 ... 600 mbar g
> -1 000 / < 2 000 mbar g
Output
Current signal
• Burden
• Auxiliary power U
B
• Current consumption I
B
Voltage signal
• Burden
• Auxiliary power U
B
• Current consumption
Characteristic
2
Piezoresistive with stainless steel
diaphragm
Burst pressure
1 bar g
1 bar g
2 bar g
2 bar g
3 bar g
4 ... 20 mA
(U
- 10 V) / 0.02 A
B
7 ... 33 V DC
(10 to 30 V for
hazardous areas)
< 20 mA
0 ... 10 VDC
> 10 kΩ
12 ... 33 VDC
< 7 mA at 10 kΩ
Linear rising
Measuring accuracy
Measurement deviation at 25 ºC (77 ºF),
Characteristic deviation, hysteresis and
repeatability included
Setting T99
Long-term drift
• Start-of-scale value
and measuring span
Ambient temperature influence
• Start-of-scale value and
measuring span
• Vibration influence ( complying
with IEC 60068-2-6)
• Auxiliary power influence
Conditions during operation
• Ambient conditions
• Ambient air temperature
– Altitude
– Relative humidity
• Storage temperature
• Degree of protection
(complying with EN 60529)
Electromagnetic compatibility≤
Construction
Weight
Process connections
Electrical connections
Material of the parts in contact with measured material
• Measuring cell
stainless steel , material no. 1.4435
• Process connection
stainless steel, material no. 1.4404 (SST 316 L)
• Sealing material
Position 15 of
order number
Viton (FPM)
A
Neoprene (CR)
B
Perbunan (NBR)
C
EPDM
D
Material of parts not in contact with the medium
• Housing
stainless steel, material no. 1.4404 (SST 316 L)
• Pin and socket
• plastic
connector housing
• CuZn, nickel-plated (plug M12)
• Cable
• PVC spec.
• typically: 0.25 % of full scale
value
• maximum: 0.5 % of full scale
value
< 0.1 s
0.25 % of full scale value/year
• 0.25 %/10 K of full-scale value
• 0.5 %/10 K of full-scale value
for a
measuring range 100 ... 400
mbar
0.005 %/g to 500 Hz in all
directions
0.005 %/V
Outdoor and indoor use
-25 ... +85 ºC (-13 ... +185 ºF)
max. 2 000 m ASL
Use an appropriate power supply
for altitudes higher than 2000
m ASL.
0 ... 100 %
-50 ... +100 ºC (-58 ... +212 ºF)
• IP65 with plug complying with
EN 175301-803-A
• IP67 with M12 plug
• IP67 with cable
• IP67 with cable fast-fit gland
• complying with EN 61326-1
• complying with EN 61326-2-3
• complying with NAMUR NE21,
only for ATEX device and max.
measured value deviation of
< 1 %
approx. 0.090 kg (0.198 lb)
Dimension drawings
• Plug complying with EN 175301-
803-A
Form A with cable inlet
M16x1.5 or ½-14NPT or Pg 11
• M12 plug
• 2- or 3-wire (0.5 mm
)
2
Cable (Ø 5.4 mm)
• Fast-fit cable gland
Media temperature
-15 ... +125 ºC (+5 ... +257°ºF)
-35 ... +100 ºC < 100 bar
(-31 ... +212 ºF; < 1 450 psi)
-20 ... +100 ºC (-4 ... +212 ºF)
-40 ... +145 ºC < 100 bar
(-40 ... +293 ºF; < 1 450 psi),
can be used for drinking water
SITRANS P210 (7MF1566)
A5E03304388-4, 06/2015