Guidance and Manufacturer's Declaration
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
This Medical-Grade UPS is intended for use in the electromagnetic environment specified below. The customer
or the user of this Medical-Grade UPS should ensure that it is used in such an environment.
Emissions Test
RF Emissions CISPR 11
RF Emissions CISPR 11
Guidance and Manufacturer's Declaration—Electromagnetic Immunity
This Medical-Grade UPS is intended for use in the electromagnetic environment specified below. The customer
or the user of this Medical-Grade UPS should ensure that it is used in such an environment.
Immunity Test
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical Fast Transient/Burst
IEC 61000-4-3
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power Frequency (50/60 Hz)
Magnetic Field
IEC 61000-4-B
Note: U T is the AC mains voltage prior to application of the test level.
18-06-068-933608.indb 12
Compliance
Group 1
SMART700HGL: Class A
SMART1200XLHGL: Class A
IEC 60601 Test Level
+ / - 6kV contact
+ / - 8kV air
+ / - 2kV for power supply lines
+ / - 1kV for input/output lines
+ / - 1kV differential mode
+ / - 2kV common mode
<5% U T
(>95% dip in U T ) for .5 cycle
40% U T
(60% dip in U T ) for 5 cycles
70% U T
(30% dip in U T ) for 25 cycles
<5% U T
(>95% dip in U T ) for 5 seconds
3 A/m
12
Electromagnetic Environment - Guidance
The Medical-Grade UPS uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and unlikely to cause any interference
in nearby electronic equipment.
The Medical-Grade UPS is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Compliance
Electromagnetic
Level
Environment - Guidance
Complies
Floors should be wood,
concrete or ceramic tile. If
floors are covered with synthetic
material, the relative humidity
should be at least 30%.
Complies
Mains power quality should be
that of a typical commercial or
hospital environment.
Complies
Mains power quality should be
that of a typical commercial or
hospital environment.
Complies
Mains power quality should be
that of a typical commercial or
hospital environment.
Complies
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
6/22/2018 4:56:31 PM