Descargar Imprimir esta página

Nikon ECLIPSE E200 Manual De Instrucciones página 71

Publicidad

2
Removal of Shipping Clamps
The stage top plate and focusing mechanism are
clamped for protection against vibration and shocks
during transportation. Remove these shipping
clamps with the hexagonal wrench provided.
• Stage Top Plate:
The top plate of the stage is retained in the
Y-axis direction by a plate fastened with two
bolts. Remove the bolts and the plate.
• Focusing Mechanism:
Remove the field lens unit to expose a plate
retaining the vertical movement of the focusing
mechanism. The plate is fastened with 3 bolts.
Remove the bolts and the plate.
3
Installing the Eyepiece Tube
Loosen the eyepiece tube clamp screw by hand and
place the eyepiece tube on the circular dovetail
mount. Tighten up the eyepiece tube clamp screw
by hand.
4
Connecting the Power Cord
Turn off the power switch of the microscope (turn to " "). Connect one end (socket) of the
supplied power cord to the AC inlet on the rear of the microscope. Connect the other end (plug)
to an AC line receptacle with the ground conductor (earth conductor). Make sure that the power
cord is securely connected.
• Note that the microscope should be installed near the AC line receptacle and the AC line
receptacle should be placed within your reach.
• Use the power cord provided. The use of other cords may damage the instrument or cause a
fire hazard.
• If using an extension cord, use only a cord that includes a protective earth (PE) wire.
This completes the assembly of the microscope standard set.
Chapter 6 Assembly
1 Assembly of Standard Set
69
Shipping
clamps
Eyepiece Tube
Clamp Screw
6

Publicidad

loading