Mechanical Problems
2
Image cannot be focused with high-power objectives.
Causes
Slide upside down.
Cover glass too thick.
High-power objective contacts slide when changed over
Causes
Slide upside down.
Cover glass too thick.
Diopter not adjusted correctly.
Difference in focal point too large when switching
Causes
Diopter not adjusted correctly.
Chapter 8 Troubleshooting Tables
2 Mechanical Problems
from low power.
from one objective to another.
81
Corrective Measures
Turn over the slide so that the cover
glass faces up.
Use a cover glass of the specified
thickness (0.17 mm).
Corrective Measures
Turn over the slide so that the cover
glass faces up.
Use a cover glass of the specified
thickness (0.17 mm).
Adjust. (P.57)
Corrective Measures
Adjust. (P.57)
8