Alarm Conditions

Priority #
Alarm
Description
1
System
Malfunction
Battery Over
Temp
2
System
Malfunction
High
Temperature
3
System
Malfunction
High
Pressure
4
No Breath
detected
5
Low Battery
Shut Down
6
Warm Up
Display
FreeStyle Comfort
Audible
More Details
Signal
1 beep
Unit will alarm when battery temperature is
every
approaching saftey shutdown threshold.
20 sec
If shutdown threshold is reached, unit will shut
down. If external power is connected, alarm
screen will display and audible alarm will
continue for 2 min.
Test with battery at room temperature. If
problem persists, contact provider.
1 beep
Unit will alarm when internal temperature is
every
approaching shutdown threshold.
20 sec
If shutdown threshold is reached, unit will shut
down and alarm screen will display and audible
alarm will continue for 2 min.
Ensure there are not blockages on any vents
of the unit. If problem persists, contact provider.
1 beep
If shutdown threshold is reached, unit will shut
every
down and alarm screen will display and audible
20 sec
alarm will continue for 2 min.
If problem persists, contact provider.
1 beep
No breath detected for 15 seconds.
every
Activated after first breath or warm up screen
30 sec
is off, whichever comes sooner.
Check cannula for kinks or leaks. If problem
persists, contact provider.
1 beep
Alarms until battery runs out of power and unit
every
shuts down.
30 sec
Charge battery.
none
Unit will start by auto-dosing until a breath is
detected or Warm up ends, whichever is sooner.
PN MN250-C4 H | User Manual